TRIGOdis
With a bigger 12" touchscreen the TRIGOdis version enables best readability from a distance in addition to high operating comfort. A solid tiltable and pivotable double holder provides optimal positioning of the central monitor and vital monitor.
The fully equipped TRIGOdis features new optional accessories such as infusion rod, multiple drawers, double holder for monitoring, lateral standard rail with secretion suction and a holder for spare cylinder.
Functions:
- 12" colour touch-screen TFT-LCD with central rotary button
- dialogue handling at the screen
- self-test and control of the single components in the standby mode
- automatic leakage test
- electronic control of the adjusted gas dosage
- electronic supervision of the pressure gas supply with display and alarm function for lower deviation in pressure
- compensation of variations in pressure supply
- detection of the system compliance
- dynamic compliance compensation
- wide adjustment range from 200 ml to 11 litres
- precise resolution and display of the scale in 20 ml steps
- automatic switchover of the scale at higher flow values
- automatic fresh gas dosing for low flow operation
- low-pressure-technology from 2 bar pressure
- display of the measured values as curves and numeric values
- configurable basic adjustments, specific for each ventilation mode
- default parameters can be defined
- integrated monitor for volume, pressure, O2, anaesthetic agents, N2O, CO2
- electronic PEEP
- for neonatal, paediatric and adult application
- error state and service information logging
- best optic and acoustic visualising of the alarms
- single vaporizer holding for different vaporizers (Halothane, Enflurane, Isoflurane, Sevoflurane, Desflurane)
- compact circle system, CO2-absorber canister is exchangeable during operation without any leakage
- RS232 interface for connection to electronic anaesthesia documentation and clinical IT-network
Security system:
- electronic O2-deficiency detection
- electronic N2O-blockage
- O2-emergency dosage (min. 4 L/min O2 flow in case of total failure)
- emergency dosage for gas supply failure: failure AIR: 100% O2, failure N2O: 100% O2, failure O2: 100% AIR
- automatic check for correct covering of the single gas supply channels every time the device is switched on
- operation is guaranteed for at least 30 minutes at power supply failure by an internal rechargeable battery
Ventilation parameters:
- CMV (volume controlled ventilation)
- PCV (pressure controlled ventilation)
- MAN./SPONT. (manual, spontaneous), spontaneous ventilation with apnea control
- SIMV (synchronous intermittend mandatory ventilation)
- MMV (mandatory minute ventilation)
- ASB (pressure support ventilation)